Laboratory analyzer water purifier BK-UP series
laboratorypolishingreverse osmosis

Laboratory analyzer water purifier - BK-UP series - Biobase - laboratory / polishing / reverse osmosis
Laboratory analyzer water purifier - BK-UP series - Biobase - laboratory / polishing / reverse osmosis
Laboratory analyzer water purifier - BK-UP series - Biobase - laboratory / polishing / reverse osmosis - image - 2
Add to favorites
Compare this product

Characteristics

Applications
for laboratory analyzers, laboratory
Purification technique
polishing, reverse osmosis, deionization, ultraviolet
Other characteristics
for ultrapure water, with accessories

Description

Product information
Brand: BIOBASE
Model: BK-UP-40L / BK-UP-60L
Origin: CHINA
Delivery time: Within 7 days
Supply: Factory direct supply

Introduction
BIOBASE BK-UP series ultrapure water purifiers deliver laboratory-grade ultrapure water with resistivity up to 18.1 MΩ·cm (@25°C). They are designed for critical workflows such as molecular biology, cell culture, analytical chemistry and instrument feed.

Features
  • Three-point water quality monitoring (feed water, RO water, UP water) with consumable life alarms.
  • 5-inch LCD for real-time status and parameters.
  • Scheduled water dispense function for routine processes.
  • High/low pressure protection and automatic shutdown on water loss.
  • Automatic start-up flushing to extend filter element life.
  • Optional pressurized water output to suit different application needs.


Purification & performance summary
Purification procedure: PF + AC + RO + AC + UV + UP + TF
UP water resistivity: up to 18.1 MΩ·cm @25°C.
RO performance: ion rejection 96%–99%; organic rejection >99%; particle and bacteria rejection >99%.

Technical specifications
  • Model: BK-UP-40L / BK-UP-60L
  • Water outlets: UP ×1, RO ×1, DI ×1
  • Output flow (25°C): 40 L/H (BK-UP-40L); 60 L/H (BK-UP-60L)
  • Purification sequence: PF → AC → RO → AC → UV → UP → TF
  • Feed requirements: inlet pressure 0.1–0.3 MPa during operation; feed water quality per GB5749; operating temperature 5–45°C
  • Purification components: PP filter ×1; granular activated carbon ×1 (post); coconut-shell activated carbon ×1; 400 GPD RO membrane ×1; 75 GPD RO membrane ×1; ultrapure columns ×4; capsule filter ×1
  • UP water quality: Resistivity 18.1 MΩ·cm @25°C; Na+ <0.1 μg/L; Cl- <0.1 μg/L; heavy metals <0.1 μg/L; Si <2 μg/L; bacteria <1 cfu/100 mL; TOC ≈20 μg/L
  • DI water quality: Resistivity 10–18.1 MΩ·cm @25°C; Absorbance (254 nm, 1 cm) ≤0.001; Soluble SiO2 ≤0.01
  • RO water quality: ion rejection 96%–99%; organic rejection >99%; particle/bacteria rejection >99%
  • Power consumption: 100 W; Power supply: AC 100–220 V, 50/60 Hz
  • Instrument size (L×W×H): 386 × 481 × 461 mm
  • Package size: Main unit 540 × 455 × 540 mm, 21.5 kg; Accessory kit 425 × 295 × 475 mm, 5 kg
  • Gross weight: 26.5 kg

Catalogs

No catalogs are available for this product.

See all of Biobase‘s catalogs

Exhibitions

Meet this supplier at the following exhibition(s):

ADLM
ADLM

28-30 Jul 2026 Anaheim (USA - California) Stand 947

  • More information
    Arablab
    Arablab

    26-28 Oct 2026 Dubai (United Arab Emirates)

  • More information

    Other Biobase products

    Other Lab Equipments

    *Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.