OverviewOperate the Integral process thermostat via the front-coloured TFT display or optionally via an intuitive touch control unit. Designed for dynamic temperature control of external circuits and compatible with a wide range of heat transfer fluids.
Key features- Process thermostat with integrated cooling system for dynamic temperature control in external circuits
- Management of heat transfer fluids with stored fluid properties
- Fully electronic continuous PID controller for internal and external control
- Low-level detection and adjustable over-temperature protection with acoustic alarm; suitable for flammable and non-flammable fluids
- Port for external Pt100; second external Pt100 possible via interface module
- Integrated adjustable bypass and large non-thermostated expansion vessel (cold-fluid layer system)
- LAUDA Variopump with 8 selectable output levels or outflow-pressure control; digital pump pressure display
- Coloured TFT display showing actual and set values and graphical temperature profile
- Programmer with 150 temperature/time segments split into up to 5 programs
- Integrated web server for browser operation in local networks; USB and Ethernet standard; optional additional interface modules (RS232/485, Profibus, analogue, contact, EtherCAT)
- SmartCool energy-saving digital cooling management with compressor on/off control
- Operates with conventional refrigerants (HFC) in accordance with EU regulation on F-gases (EU) 573/2024
- SelfCheck assistant for system diagnosis and PowerAdapt for optimised use of available supply power
Working range- Min. working temperature: -90 °C
- Max. working temperature: 220 °C
- Temperature stability: 0.05 ± K
Technical data (according to DIN 12876)Working temperature range: -90 ... 220 °C
Ambient temperature range: 5 ... 40 °C
Temperature stability: 0.05 ± K
Heater power max.: 12 kW
Max. power consumption: 19 kW
Current consumption: 30 A
Max. discharge pressure: 3.1 bar
Max. pump flow (pressure): 65 L/min
Inlet/outlet connection thread (outside): M38 x 1.5 / Hose size: 1″
Filling volume: min. 10 L — max. 30.6 L
Cooling water connection (thread outside): 3/4 ″; recommended cooling water temp.: 15 °C; max 30 °C; flow 26 L/min; pressure diff. min. 0.8 bar — max 10 bar
Dimensions (W×D×H): 760 × 650 × 1605 mm
Weight: 356 kg
Noise level: 65 dB(A)
Power supply: 400 V; 3/PE; 50 Hz & 460 V; 3/PE; 60 Hz
Power plug: Power cord with plug (IEC 60309, 5-pin, CEE, red, 32 A)
Cooling capacity (selection)200 °C | Pump level 8 | Thermal oil | 18.5 kW (50/60 Hz)
100 °C | Pump level 8 | Thermal oil | 18.5 kW (50/60 Hz)
20 °C | Pump level 8 | Ethanol | 18.5 kW (50/60 Hz)
10 °C | Pump level 8 | Ethanol | 15 kW (50 Hz) / 16 kW (60 Hz)
0 °C | Pump level 8 | Ethanol | 11.5 kW (50 Hz) / 12.2 kW (60 Hz)
-10 °C | Pump level 8 | Ethanol | 8.7 kW (50 Hz) / 9 kW (60 Hz)
-20 °C | Pump level 4 | Ethanol | 8.5 kW (50 Hz) / 9 kW (60 Hz)
-30 °C | Pump level 4 | Ethanol | 8.5 kW (50 Hz) / 9 kW (60 Hz)
-40 °C | Pump level 4 | Ethanol | 7.5 kW (50 Hz) / 8.5 kW (60 Hz)
-50 °C | Pump level 4 | Ethanol | 6 kW (50 Hz) / 7 kW (60 Hz)
-60 °C | Pump level 4 | Ethanol | 4 kW (50 Hz) / 5.2 kW (60 Hz)
-70 °C | Pump level 4 | Ethanol | 2.2 kW (50 Hz) / 3.4 kW (60 Hz)
-80 °C | Pump level 4 | Ethanol | 0.9 kW (50 Hz) / 1.5 kW (60 Hz)
-90 °C | Pump level 2 | Ethanol | 0.35 kW (50 Hz) / 0.5 kW (60 Hz)
Technical specifications- Model: Integral IN 1590 XTW
- Working temperature range: -90 to 220 °C
- Temperature stability: 0.05 ± K
- Heater power (max): 12 kW
- Max. power consumption: 19 kW
- Current consumption: 30 A
- Max. pump flow pressure: 65 L/min
- Filling volume: min. 10 L — max. 30.6 L
- Interfaces: USB and Ethernet standard; optional RS232/485, Profibus, analogue, contact or EtherCAT modules
- Integrated web server and programmer (150 segments / 5 programs)
- Dimensions (W×D×H): 760 × 650 × 1605 mm; Weight: 356 kg